{"id":612,"date":"2023-09-13T11:03:00","date_gmt":"2023-09-13T11:03:00","guid":{"rendered":"https:\/\/quantadt.wpengine.com\/us\/?post_type=faq&#038;p=612"},"modified":"2025-12-04T15:39:05","modified_gmt":"2025-12-04T15:39:05","slug":"system-flow-balance-optimization-and-fluid-removal-accuracy","status":"publish","type":"faq","link":"https:\/\/quantadt.com\/us\/faq\/system-flow-balance-optimization-and-fluid-removal-accuracy\/","title":{"rendered":"System flow balance optimization and fluid removal accuracy"},"content":{"rendered":"\n<p class=\"wp-block-paragraph\">The use of passive and active control allows errors to be minimized across a wider dynamic range of conditions. For the Quanta System (previously known as the SC+), the average flow balance error was 1 mL\/hr with an SD of 19 mL\/hr, and with ultrafiltration it was 13 mL\/hr and an SD of 20 mL\/hr across all conditions. This paper demonstrates that the hemodialysis system, a small, simple, and versatile CE marked device, operates within the limits required by international standards across a wide range of experimental conditions.<\/p>\n","protected":false},"template":"","class_list":["post-612","faq","type-faq","status-publish","hentry"],"acf":[],"_links":{"self":[{"href":"https:\/\/quantadt.com\/us\/wp-json\/wp\/v2\/faq\/612","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/quantadt.com\/us\/wp-json\/wp\/v2\/faq"}],"about":[{"href":"https:\/\/quantadt.com\/us\/wp-json\/wp\/v2\/types\/faq"}],"wp:attachment":[{"href":"https:\/\/quantadt.com\/us\/wp-json\/wp\/v2\/media?parent=612"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}